YSI 6 series Manual de usuario Pagina 35

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 39
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 34
CALIBRATION PROCEDURE FOR DISCRETE SAMPLING
The YSI Rapid Pulse and ROX DO sensors can be calibrated in saturated air or saturated water; both
will give the same results when done correctly. For these deployments we recommend that you calibrate
in saturated water. Saturated water will allow you to check the performance of the other sensors and
eliminate any chance of temperature-related calibration errors. All that is required is a clean five-gallon
bucket, a good quality aquarium air pump, and two air stones.
1. Fill pail to at least 75% capacity with room-temperature tap water.
2. Connect air stone to pump and sparge water for at least one hour before calibration.
3. Place the sonde in the water and run in “Discrete Mode” at a one second rate.
4. Wait 10 minutes to ensure temperature stability.
Note: The Rapid Pulse DO sensor should read very close to the saturated air reading of your
previous air calibration. If it reads more than 2% higher than the reading in saturated air, then
you could have a Rapid Pulse membrane problem. Check for leaks, punctures, or a loose-fitting
membrane retainer o-ring before you calibrate.
5. Calibrate both the Rapid Pulse and ROX DO sensors. Record the barometric pressure at the time
of calibration.
Note: Avoid having air bubbles collect on either membrane, during the calibration.
6. Barometric pressure ______________mmHg
Note: pressure divided by 7.6 = calculated DO
Example: 760 ÷ 7.6 = 100 %
7. ROX Output ________________% _____________mg/l ± 1%
8. RP Output _________________ % _____________mg/l ± 2%
9. Temperature ____________°C, Salinity _____________ppt
10. DO NOT recalibrate the ROX DO sensor in the field!
11. Do your post-calibration when you return in the same saturated water bath.
Stability and agreement must occur in this controlled environment or good field measurements will not
be made. The ROX and Rapid Pulse membranes must be kept in a saturated air or water during
transportation to the field and in between readings.
Notes:
(continued)
32
Vista de pagina 34
1 2 ... 30 31 32 33 34 35 36 37 38 39

Comentarios a estos manuales

Sin comentarios