Calibration,Maintenance
&Troubleshooting Tips
For YSI 6-Series Sondes & Sensors
Vista de pagina 0
1 2 3 4 5 6 ... 38 39

Indice de contenidos

Pagina 1 - &Troubleshooting Tips

Calibration,Maintenance&Troubleshooting TipsFor YSI 6-Series Sondes & Sensors

Pagina 2 - Contents

Note: the sonde will only read 100% saturation if the barometric pressure is 760 mmHg. To determine the actual DO in %, divide the barometer reading b

Pagina 3 - CALIBRATION WORK SHEET

TURBIDITY6136PROBECALIBRATIONTIPS Note: The calibration of all YSI turbidity sensors must be done with either YSI distributed standards, Hach S

Pagina 4

4. Always ensure that the submerged parts of the sonde and wipers are clean before beginning any turbidity calibration. Remove any brush wipers and r

Pagina 5

OPTIMIZINGTHEPERFORMANCEOFTHE6136TURBIDITYPROBEFORLOW‐TURBIDITY&FOULINGENVIRONMENTS The following instructions will improve the meas

Pagina 6 - ORP/REDOXCALIBRATIONTIPS

INCREASE THE SENSOR SAMPLING PERIOD 1. Go to the sonde’s Advanced Menu, then to Data Filter. Verify that the Data Filter is enabled. 2. Go to th

Pagina 7

6820 /6920 (Single Optical Port): Calibration of these sondes can be done in the standard YSI calibration cup. The correct calibration cup measures 7.

Pagina 8

The only true way to determine if your zero standard is being contaminated is to have the zero solution analyzed in a instrument like the Hach 2100AN

Pagina 9

CHLOROPHYLLCALIBRATIONTIPS BEFORE YOU BEGIN: Make sure that the sonde firmware is version 3.08 or higher. Upgrade at www.ysi.com SET UP & OPER

Pagina 10

4. The following table is for the Rhodamine WT solution. Instructions for preparation can be found in the Principles of Operations (Section 5) of yo

Pagina 11

CONDUCTIVITY&TEMPERATURETROUBLESHOOTING 6560 TEMPERATURE & CONDUCTIVITY SENSOR The temperature sensor should be tested from time to time

Pagina 12

Contents Calibration Work Sheet i Conductivity 1 pH 2 ORP 3 Dissolved Oxygen - Rapid Pulse 6562

Pagina 13

pHTROUBLESHOOTING pH probes typically have a shorter life expectancy than the other sensors on your sonde. They can be expected to last at least on

Pagina 14

OUT OF RANGE The out-of-range message is usually confirming that the pH probe has a bad slope and needs to be cleaned and reconditioned. If this reco

Pagina 15

RAPIDPULSEDISSOLVEDOXYGENTROUBLESHOOTING The Rapid Pulse DO sensor used in many 6-Series instruments can be quickly tested when a problem is susp

Pagina 16

HIGH OR LOW DO GAIN (RANGE 0.7 TO 1.4) The DO gain number is a multiplier. It is usually out of tolerance when the DO charge or the High/Low warm-up

Pagina 17 - CHLOROPHYLLCALIBRATIONTIPS

ROX™OPTICALDISSOLVEDOXYGENTROUBLESHOOTING POST-CALIBRATION IS LOW Verify that the ROX is in a saturated environment. If yes, then remove the wip

Pagina 18

CHECK THE ROX MEMBRANE SURFACE The black painted surface covering the ROX membrane is necessary for the probe to function properly in directly focuse

Pagina 19

DEPTHTROUBLESHOOTING NON-VENTED DEPTH On occasion the depth sensor can be affected by biological fouling that grows in the water passage tube. The

Pagina 20

TURBIDITYTROUBLESHOOTINGBEFORE YOU BEGIN: Update the sonde’s firmware to the version 3.08 or higher. With this update you MUST now use a black wipe

Pagina 21

PORT AND PROBE CLEANING PROCEDURE To remove contamination -- which can include grease, saltwater, electrolyte, and calibration reagents -- the probes

Pagina 22

SONDE CONNECTOR & FIELD CABLE SERVICING PROCEDURE BEFORE YOU BEGIN - Clean and rinse the connectors with fresh water before you disassemble them.

Pagina 23

CALIBRATION WORK SHEET Date of Calibration: ___________________ Sonde ID: ________________________ Te

Pagina 24

(Above Left) Example of a correctly mated connector. Note: There are typically only two exposed threads when the connector is fully inserted and o-r

Pagina 25

UNATTENDEDMONITORING “MAKINGGOODDOMEASUREMENTS” It is recommended that you take the following steps prior to d

Pagina 26 - DEPTHTROUBLESHOOTING

3. At the recovery, the process is repeated. The sonde, when recovered, would be externally wiped clean (not the sensors) and then wrapped again in

Pagina 27 - TURBIDITYTROUBLESHOOTING

DISCRETESAMPLING“MAKINGGOODDOMEASUREMENTSWITHTHERAPIDPULSEDOPROBE” Preparing a sonde for use in a spot sampling/profiling application is

Pagina 28

COMPARINGROX&RAPIDPULSESENSORS The instructions and QA recommendations on these pages will be used to set up and calibrate YSI 6- Series so

Pagina 29

CALIBRATION PROCEDURE FOR DISCRETE SAMPLING The YSI Rapid Pulse and ROX DO sensors can be calibrated in saturated air or saturated water; both will g

Pagina 30

UNATTENDED DEPLOYMENTS 1. Install the 6150 ROX optical DO sensor into a spare optical sensor port. 2. Connect the sonde to a PC. At the command pr

Pagina 31

CALIBRATION PROCEDURE FOR UNATTENDED DEPLOYMENTS The YSI Rapid Pulse and ROX DO sensors can be calibrated in saturated air or saturated water; both w

Pagina 32

ECOWATCHPAGEWIDTHEXPANSIONPROCEDURE ML 1/19/2009 This procedure prevents data wrap-around when calibrating or using the Discrete Run mode in YSI

Pagina 33 - DOPROBE”

YSIIncorporated1700/1725 Brannum Ln Yellow Springs, Ohio 45387 USA Environmental Technical Support : 800-897-

Pagina 34

CONDUCTIVITYCALIBRATIONTIPS BEFORE YOU BEGIN: Verify the accuracy of your temperature probe with a traceable thermometer or other temperature refer

Pagina 35

pHCALIBRATIONTIPS 1. Go to the sonde’s Report Menu and enable the pH mv output. This will allow the sonde to display the millivolt output from th

Pagina 36

ORP/REDOXCALIBRATIONTIPS 1. The ORP and pH sensors are combined on all current YSI sondes. You must calibrate the pH sensor first and insure that

Pagina 37

DISSOLVEDOXYGENCALIBRATIONTIPS UNATTENDED & SAMPLING DO SENSOR PREPARATION 1. Inspect the DO probe anodes; recondition using the 6035 recond

Pagina 38 - ML 1/19/2009

6. The probe is now ready to be calibrated. Set the Auto Sleep RS-232 for the intended application: ON for UNATTENDED STUDIES and OFF for SAMPLING

Pagina 39

ROX™OPTICALDISSOLVEDOXYGENCALIBRATIONTIPS ROX FACTS • The ROX Optical Dissolved Oxygen sensor does not require any special sonde setup or bur