NLN Normal, linear logged scan
NLA Normal, linear average logged scan
NLG Normal, logarithmic logged scan
NEB Normal, event based logged scan
AHI Alarm, logged scan above high alarm limit
ALO Alarm, logged scan below low alarm limit
WHI Warn, logged scan above high warn limit
WLO Warn, logged scan below low warn limi
Table 1, Broadcast ID code description
3.4.3 CHANGE BATTERY
The Level Scout uses two AA alkaline batteries for power. When these batteries are changed, the transducer’s internal battery
condition must be reset so that the percentage of battery charge remaining is accurately reported. This parameter should only be
modified if the transducer’s batteries are replaced. Two types of batteries may be selected as having been installed. YSI recommends
only using AA alkaline batteries in the Level Scout. If the incorrect battery type is selected, the available power indicator will be
inaccurate. Normally, you cannot input a value in the ‘Capacity in mAhr’ box. However, if you select ‘Other’ you may input the
correct milliamp-hours listed for the battery to receive accurate battery capacity readings.
Click the Change Battery button (Figure 3.27) to access the ‘Change Batteries’ window (Figure 3.28). Select the battery type and click
OK to save the change to the battery, then confirm by clicking OK again in the warning window.
Due to the inherent explosion risks of lithium batteries when exposed to
water, YSI does not recommend using lithium batteries in the Level Scout.
Warnin
Figure 3.27
Figure 3.28
YSI Incorporated Data Scout 19